We can calibrate your OTDR at two service price levels: ISO9001 or ISO/IEC 17025
- We use an automated connector inspection probe to check the optical connector end face quality against IEC61300-3-35. If we find a performance problem with the received instrument, we will let you know.
- We test accuracy of measurement of: length, attenuation, wavelength, dynamic range, ORL
- We have NATA traceability for OTDR power, attenuation, ORL & wavelength parameters
- We have an in-house fiber-length test standard, approx 10 Km
- Test methods comply with the latest OTDR calibration standards EN61746-1-2011+AC-2014 and EN61746-2-2011+AC-2014
- Make use of our long experience to also calibrate any source / meter ports